How To Set The Temperature Of The Constant Temperature Shaker
Nov 16, 2018
Leave a message
Turn on the power switch and set the constant temperature:
(1) Put the control small switch in the “Setting” section. At this time, the temperature displayed on the display is the set temperature. Adjust the knob and set it to the temperature you need to work. (The operating temperature you set should be higher than the ambient temperature. At this point, the machine starts to heat up, the yellow indicator light is on, otherwise the machine does not work)
(2) Put the small switch of the control part on the “measurement” end. At this time, the temperature displayed on the display screen is the actual temperature of the air in the test chamber. As the temperature inside the box changes, the displayed number will change accordingly.
(3) When heated to the temperature you need, the heating will stop automatically and the green indicator light will be on; when the heat in the test chamber is dissipated below the temperature you set, a new round of heating will start again.
Application of shaker in mammalian cell culture
Culture conditions: susceptible to mechanical damage, low speed, 5% CO2 conditions.
Culture characteristics: Rmp is less than 100 37 ° C 5% Co2

