MDS-2014DS Dissolution Drug Tester
video

MDS-2014DS Dissolution Drug Tester

1.※The dissolution vessel is designed with embedded fixing hoop to achieve better centering and verticality of the dissolution vessel. Indicator strips and diagonal fixtures are designed on the cup rim to achieve higher reproducibility. Available in 1000mL and 250mL specifications, and the dissolution vessel can be either transparent or low actinic red for photosensitive substances.dissolution drug tester
Send Inquiry

Description

Technical Parameters

Model No:MDS-2014DS dissolution drug tester

1.Parameters

1. ※The dissolution vessel is designed with embedded fixing hoop to achieve better centering and verticality of the dissolution vessel. Indicator strips and diagonal fixtures are designed on the cup rim to achieve higher reproducibility. Available in 1000mL and 250mL specifications, and the dissolution vessel can be either transparent or low actinic red for photosensitive substances.

2. The V-cup cover is designed to fit well with the dissolution vessel, which can reduce solvent evaporation.

3. an automatic dosing module that avoids tablet sticking, enabling simultaneous and sequential dosing.

4. ※The one-piece injection-molded water bath box with quick adapter as standard, there is no bonding in the box, and there is no dead angle area such as right angle.

5. ★ It adopts segmented 316L stainless steel propeller shaft design. When replacing, only the lower propeller shaft is required to be replaced, and the whole root is not disassembled, which reduces the workload of re-setting the height after replacement.

6. ★ The machine head adopts multi-point stainless steel guide rod to ensure the stability of the machine head and effectively prolong the stable time of mechanical performance.

7. ※ A color touch screen of no less than 10 inches is used for human-computer interaction , which can realize the setting, storage and operation of no less than 100 methods. With screen lock security function, it also provides Chinese and English bilingual operation interface.

MDS-2014

8. Provides three levels of administrative privileges compliant with audit trail requirements to control access and prevent unnecessary changes to instrument settings or methods.

9. The instrument is equipped with an independently controlled heating cycle system with an unattended reservation function.

10. The instrument has eight independent temperature probes, and the temperature of the eight cups of solvent is independently monitored and displayed in real time. The temperature control method should adopt PID+fuzzy control, the temperature control accuracy should not be lower than ±0.2℃, and the temperature resolution should reach 0.1℃.

11. The instrument should have a built-in printer, which can print and save experimental data in real time.

12. The instrument has built-in photosensitive lighting, and with the brown dissolution vessel, light-proof experiments can be realized.

13. ※The instrument can store 100 experimental methods, and the instrument methods can be imported and exported through one-key USB operation, avoiding the tedious steps of method transfer between different instruments.

14. Manufacturers provide standard-compliant AIQ certification services, mechanical verification services, and technical and application-related support services such as instrument training and installation on-site training.

15. ※The sampler is equipped with an automatic door opening device, which can realize one-key entry and exit of the warehouse.

16. It can automatically calculate the sampling height, and at the sampling time, it can accurately sample the position in line with the Pharmacopoeia according to the volume of the solvent;

17. There should be a secondary filter interface to make the impurities filter more fully.

18. ★ The instrument adopts imported syringes, which can realize the steps of automatic rinsing, filling, waste drop disposal, sampling, replenishment, and emptying, etc., to ensure the sampling accuracy, and to ensure that the pipeline residue is as little as possible.

MDS-2014DS

19. The instrument adopts imported PTFE micro-transparent pipeline, which has anti-corrosion and anti-adsorption properties.

20. ★ Continuous sampling, dual sample sampling, automatic sample dilution and other functions, the shortest sampling interval is 2 minutes

21. The pipeline and sampling pump are automatically rinsed before sampling to ensure the accuracy of sampling samples.

22. After the experiment is completed, one-click automatic cleaning of the liquid tank and pipeline ensures no cross-contamination.

23. The product can provide professional third-party testing report on product performance.

 

 

dissoultion with sampler

 

Hot Tags: MDS-2014DS Dissolution Drug Tester, China, factory, suppliers, wholesale, customized, buy, cheap, bulk, price list, for sale

1. Dissolution Apparatus

project

index

Instrument performance index

Water bath heating range: room temperature -45. 0 ℃

Temperature Resolution: 0.1 ℃

Temperature control accuracy: ±0.2 ℃

Stirring speed range: 25.0-250.0 RPM

Speed Resolution: 0.1 RPM

Speed Accuracy: ±0.5 RPM

Instrument mechanical indicators

Basket (paddle) shaft swing amplitude: ≤0.5mm

Basket swing range: ≤ 1.0 mm

rotating shaft and the dissolution vessel shaft: ≤ ± 1.0mm

Dissolution apparatus horizontality: 0.5°

Basket (paddle) axis verticality: 0.5°

Dissolution vessel verticality: 1.0°

Instrument function

Compatible with basket method, paddle method, small cup method, paddle butterfly method, rotating drum method

Number of dissolution vessels: 6/8 cups

Segmented shaft design, no need to manually re-adjust the height of the positioning basket (paddle) when the dissolution method is switched.

Automatic simultaneous dosing/sequential dosing

Photosensitive lighting: white light/red light three-level lighting

Can store 100 experimental methods, and can import and export USB

It can store 39 login accounts and passwords, and the operation authority can be controlled in 3 levels

With operation record printing and log recording function, and can audit trail

input power

220Vac ± 10% 1300W _

Overall size

670*565*640(W×D×H)

Total Weight

6 5kg _

Operating temperature

(5~40) ℃

storage temperature

(-20~60) ℃

Working humidity

(2 0 ~ 80) % RH

Storage humidity

(5~95) % RH

2. Sampler

project

index

Performance parameters

Single sampling volume: 0.2-15mL (single); 0.4-30mL (double)

Sampling times: 1-16 times (single); 1-8 times (double)

Sampling accuracy: ≤0.8% (10mL sampling)

Sampling rack: 2ml liquid phase injection bottle * 128 or 1 5 ml test tube * 128

Instrument function

Fully automatic rinsing, filling, dripping, sampling, replenishment, dilution, emptying, cleaning

Precise sampling by syringe pump, suction and stop time can be set

input power

220Vac ± 10% 200W

Overall size

515*660*460mm(W×D×H)

Total Weight

35Kg

Operating temperature

(5~40) ℃

storage temperature

(-20~60) ℃

Working humidity

(20~80) % RH

Storage humidity

(5~95) % RH